In this study we show the application of a nondestructive process
analytical technology to separate two populations of enteric coated
capsules that contained minitablets. In the course of a small, phase 1
clinical trial investigation it was deemed necessary to have two different
placebo capsule formulations that were matched in appearance.
These capsule batches were manually filled wherein subsequent processing
required that they be specifically identified and accordingly
classified (i.e., sorted). The sorting was based on a near-infrared (NIR)
spectral differentiation between the capsule fill of the two formulations.
Both capsule lots appear identical as enteric coated: white (titanium
dioxide) hard gelatin capsules having an orange band. One lot
contains pure microcrystalline cellulose (MCC) minitablets while the
other lot contains minitablets composed principally of sodium caprate
(C10) with PEG 3350 added as a binder. The enteric coating used on
both lots is composed of methacrylate polymers with plasticizers at a
cured coating level of approximately 14 mg solids per cm2 of capsule
surface area. In order to facilitate a streamlined manufacturing schedule,
it was necessary to rapidly develop and GMP qualify a technically
feasible, nondestructive capsule identification and sorting process.
Near-infrared (NIR) spectroscopy has successfully been applied
to the noninvasive and nondestructive differentiation of various capsules,
tablets, and drugs in solution [1-5]. While current scholarship
suggests that NIR is fairly well established for this purpose, previous
studies focus on the utility of NIR for the quantification and identification
of capsules filled with powders and solutions. In the current case,
the formulations in the capsules (methacrylate coated, orange-banded,
hard gelatin) were not powders or solutions, but were minitablets. Due
to the large particle size (2 mm) and random orientation of the
minitablets in the capsules, the NIR absorbance spectra demonstrated
more variation than a uniformly filled powder capsule. The combined
effect of the enteric-coated hard gelatin exterior, and the random distribution
of the minitablet interior, made analysis of these capsules potentially
more challenging than those in previous experiments. We undertook
a combination of selective wavelength analysis along with
chemometric principles to develop a non-destructive method for distinguishing
between these two capsules. The method was subjected to
validation testing using a calibration set and was ultimately used to sort
through two populations of capsules totaling 3407 units using only two
near-IR wavelengths provided by interference filters from a tungstenhalogen
source. Such a simple instrument can be easily manufactured
and widely deployed as a process sensor.
The near-infrared region of the electromagnetic (EM) spectrum
covers wavelengths from 750 nm to 3000 nm. This region consists of
broad, overlapping peaks that result from overtone bands, combination
bands, and difference bands from molecular vibration of CH, NH, OH,
and SH bonds [6]. Many factors contribute to the variations in NIR
spectra, such as detector noise, environmental conditions, and
different
sample preparation. Often the largest variations come from the
constituent
or concentration differences. Using this knowledge, chemometric
techniques such as principal axis transformation can be used to
interpret complex overlapping spectra by placing the original spectral
variables into a new, smaller coordinate axis system [7]. Calculation
of
the principal components is accomplished by a singular value
decomposition
of matrix A according to A=USV, where A is the matrix of
original spectra, U is the matrix of eigenvalues (scores), S is a diagonal
matrix of singular values, and V is the matrix of eigenvectors (loadings).
The first principal component may capture 80% or more of the
total variance. Many of the eigenvalues model only noise, therefore
only those that contain a significant proportion of the variation with
analytical signal are used in calibration and evaluation of samples.
Principal components can be used to illustrate the separation of
two groups. For a more quantitative result, the bootstrap error-adjusted
single sample technique (BEST) was applied to the calibration data.
The BEST algorithm begins by encoding the intensity on each wavelength
as a separate dimension, thus reducing each spectrum to a single
point in multidimensional space. Population P is an m x n matrix
in hyperspace R whose rows are the individual samples and the
columns are the frequencies [8]. BEST considers each wavelength
from a spectrum of n wavelengths to be taken as a separate dimension,
such that each spectrum is reduced to a single point in n-dimensional
hyperspace [3]. P* is a discrete realization of P based on a calibration
set T of the same dimensions as P*. This realization is chosen one
time from P to approximate all possible sample variations present in P.
P* has parameters B and C, where C = E(P) and B is the Monte Carlo
approximation to the bootstrap distribution. The expectation value,
E(P), is the center of P, and is a row vector with the same number
of rows as there are columns in vector P. New test spectra X are projected
into hyperspace R containing B, rows of B are mapped onto a
vector connecting C and X. C and X have the same dimensions. The
integral over R is calculated from the center of P in all directions. A
skew-adjusted standard deviation (SD) is based on the comparison of
the expectation value C=E(P) and CT), the median of T in
hyperspace projected on the hyperline connecting C and X. The result
is an asymmetric SD that provides two measures of the SD along the
hyperline connecting C and X. Equation 1 defines the SD in the direction
of X, and Equation 2 defines the SD in the opposite direction.
Skew adjusted SDs can be used to calculate mean distances between
spectra of different samples.

The first objective was to prove that full NIR spectra from
1100-2500 nm (in 2 nm steps with 10 nm bandpass) could be used to
separate
the MCC and C10 capsules. Initial experiments involved only
three MCC capsules and three C10 capsules. Spectra were collected
with a scanning monochromator instrument. Each capsule was
scanned three times for a total of 18 spectra (Figure 1).

The capsules were scanned in random order and rotated following
each scan to average possible sample variations due to inconsistencies
in the enteric coating, gelatin layer and the orientation of the
minitablets. Scans were collected inside of an instrument drawer to
eliminate room noise and external interferences. Data were multiplicative
scatter-corrected to eliminate baseline variations [9], and second
derivatives [10] were plotted to find the regions in the spectrum where
the most variation was apparent (Figure 2).

Principal components (PCs) were calculated and the PCs with
the largest contribution to variation were plotted in two and three
dimensions to visualize the differences between capsule groups
(Figures 3 and 4) [7]. BEST standard deviations and cross-validation
standard deviations (CV-SD) provided a quantitative measure of group
separability [3,8].


In order to estimate the limits of detection for the NIR, a cluster
translation procedure was performed [11]. NIR spectra from the two
capsule populations, P1 and P2 are expressed in m x n matrices, where n
is the number of wavelengths and m is the number of spectra. The
columns of the matrices are averaged by Equations 3 and 4, giving two
1 x n vectors.
A difference spectrum X is calculated from P2ave – P1ave. One population
was spatially translated toward the other, PAdjusted = y*X+P2,
where y starts at zero, increasing in increments of 0.01 until P1 and
PAdjusted are inseparable. It is assumed that the two capsule groups represent
the possible variations in the pure component spectra (MCC and
C10), and that all points on the hyperline connecting the centers of the
two population distributions correspond to mixtures of the two components
because the Beer-Lambert Law holds. For example, when one
population is translated one-half the distance toward the other, that population
corresponds to a 50/50 mixture of MCC and C10. The maximum
distance the two groups can move toward each other while maintaining
statistical separation determines the minimum quantities of
MCC and C10 that can be detected in each other.

Although the overall goal of this project was to develop a
process
analytical method sensitive enough to assign the MCC and C10 capsules
to their respective groups accurately, it was also imperative that the
project be completed as quickly as possible. This time constraint
required a modification of the full spectrum approach. Experiments
were conducted to accomplish the separation with as few wavelengths
as possible, yet still with adequate statistical assurance of
specificity.
Using the same six capsules scanned above at 701 wavelengths, new
spectra were collected with a 19-wavelength filter wheel NIR
spectrometer
between 1445 and 2348 nm.
Capsules were placed in the conical reflective cup oriented with
the thicker cap end down, body end up, and held in place with a steel
rod (Figure 5). The conical reflecting cup is designed such that when
a sample capsule is placed along the axis of radial symmetry of the
cone, specular reflection at the detector is minimized while diffuse
reflectance is maximized [2].

All radiation that follows a path parallel to the incident beam and
perpendicular to the base of the conical reflector is reflected and collimated
back toward the source. This radiation is predominantly specular
reflectance and contains little information about the capsule fill. In
the same fashion, a small amount of radiation that passes through the
capsule but does not scatter is also returned to the source. Therefore,
the majority of the radiation reaching the detector via the integrating
sphere is scattered by the contents of the capsule. The amount of radiation
that reaches any given location on the capsule is directly proportional
to the curved surface area of the frustum (the conic section
defined by two parallel lines from the light source, and a plane parallel
to the reflector base connecting the two lines) in which it lies.
Therefore, more light reaches the top of the capsule than the bottom of
the capsule because there is more curved surface area at the top of the
reflector than at the bottom.
The curved surface area is given by πs(r1+r2) where r1 and r2 are
the radii of the base and top of a circular frustum, s is the length
between the top and bottom measured along the surface of the cone. If
the detector collected scattered light from cross-sections of the capsule,
there would be a different intensity value for each cross-section.
However, the detector uses an integrating sphere, which collects all of
the scattered light from the entire capsule. This configuration eliminates
the concern of uneven illumination along the capsule.
In order to best accommodate the capsules, the spectrometer was
inverted, and an instrument drawer was fashioned for consistent and
reproducible sample loading. Each scan took approximately two minutes
to collect the full 19 wavelengths. Capsules were scanned in random
order and rotated to average sample positioning variations.
Principal components, intercapsule BEST standard deviations (SDs),
and intracapsule cross-validation standard deviations were calculated
from the resulting spectra. Spectra for the two groups of capsules were
very different, allowing a visual inspection of the spectra to
sufficiently
identify the more distinguishing wavelengths. Standard deviations
and cross-validation-SDs were calculated from four selected
wavelengths.
To operate as quickly as possible while maintaining the highest
level of capsule classification accuracy, the two most distinguishing
wavelengths were selected by visual inspection. From these wavelengths,
ratios were calculated and plotted to prove that the two wavelength
approach was sufficient to justify exploration of a larger validation
set.
The filter wheel spectrometer was installed in the GMP facility
in the University of Kentucky Center for Pharmaceutical Science &
Technology, and turned on for the remainder of the procedure to eliminate
detector drift due to thermal variations. Apolystyrene calibration
film standard fitted to a conical reflective cup was scanned 50 times to
capture all possible sample variations. A ratio was calculated from the
signal intensity at 1734 nm and 1445 nm for each scan of polystyrene
standards, and confidence limits were constructed around the mean at
± six standard deviations, 1.0245 ± 0.0118. The choice of a 12 standard
deviation acceptance range was made to ensure essentially 100%
confidence limits on classifications for the set of 3407 capsules to be
scanned. To prove that the instrument response was the same from day
to day, these ratios were projected on top of the predefined constructed
confidence limits.
The same approach was used to identify the different capsules.
A calibration set consisting of 50 MCC and 50 C10 capsules was
scanned. Wavelength ratios were calculated and confidence limits
were constructed at the mean ± six standard deviations for each of the
two capsule formulations. These limits were defined as the acceptance
criteria boundaries for the determination of which group the incoming
capsule belonged (Figure 6).

At the start of each day of data collection, the polystyrene calibration
standard was scanned to prove the instrument was performing reproducibly.
Capsules were sequentially scanned in groups of 20; a total of
3407 capsules were scanned. An algorithm was written to automatically
calculate and display the wavelength ratios on top of the predefined confidence
intervals (Figure 7). Capsules were sorted according to where their
NIR ratios projected relative to the calibration experiments.

Figure 8 illustrates the two wavelengths that were selected to distinguish
between the two groups of capsules. The inert MCC has a relatively
flat spectrum, while C10 has a very steep spectrum at these wavelengths.
It is this spectral feature that allowed the simple calculation of a ratio from
two wavelengths to distinguish between the two groups.

Cross-validation results and the BEST standard deviations
between
capsules are reported in Table 1 for the full spectrum NIR
measurements,
for four wavelengths, and for two wavelengths. Note that the standard
deviations are the same magnitude with two wavelengths as with the full
spectra, suggesting that the basis of selectivity is improved by
focusing on
those wavelengths which can unequivocally distinguish between groups.
This approach justifies and allows for a much smaller data set with
which
to base the classification upon. Illustrated in Figure 9 are the ratios
(1734/1445 nm) calculated from all capsules projected into the
confidence
intervals determined during the calibration stage of the experiment.


This experiment resulted in greater than 99.71% successful capsule
identification. Of the 3407 capsules scanned, ten capsules projected directly
on or just outside their respective decision boundaries. These outliers
were scanned a second time at the end of the experiment, and their twowavelength
ratios projected inside their respective confidence intervals.
These results proved that the unclassified capsules were a result of erroneous
manual sample loading, and further demonstrated the consistency
and integrity of both the method of analysis as well as the capsules themselves.
Detection limits were estimated by the spatial cluster
translation
experiment described in the Methods above. When using MCC capsule
spectra as the calibration set and C10 capsule spectra as the test set,
C10
capsules could be spatially translated 92% of the distance across space
before the populations were inseparable. When using C10 as the
calibration
set and MCC as the test set, MCC was spatially translated 89% of the
distance across space before the populations were inseparable. This
demonstrates that NIR is capable of nondestructively identifying
mixtures
of C10/MCC in the capsules down to approximately ten percent of each
one in the other. It is apparent that the NIR detection limits far
exceed what
is necessary to distinguish between MCC and C10 capsules.
This study presents an effective application of NIR spectroscopy to
the noninvasive and nondestructive classification of MCC and C10
minitablets contained in enteric-coated (methacrylate polymer) hard gelatin
capsules. The experiment was conducted in a GMP facility and is relevant
to PAT. It was a very rapid method with data collection times of only
10 seconds, and the two wavelength approach gave unequivocal separation
between the two capsules. The entire process, from conception to completion,
required only 20 days. Of the 3407 capsules scanned, greater than
99.71% projected into their respective confidence intervals. Of the ten capsules
that initially failed to validate, it was proven that human sample loading
errors were responsible, suggesting that an automated version of the
same experiment could have classified 100% of the capsules correctly.
One of the primary goals of the FDA PAT initiative is to increase automation
to reduce human error. This experiment provides an example of the
benefits of automation, as well as of the utility of a method for real-time
characterization and release of individual drug products.
New measurement, control and information technologies are needed
in PAT to predict, control and assure product quality and performance.
Using an appropriate PATsensor, product quality attributes can be accurately
and reliably predicted over the design space established for the materials
used, the process parameters, and the environmental and other conditions.
Atwo-wavelength near-IR sensor for PAT will be rugged, inexpensive, and
simple to construct using two interference filters and a detector (e.g., PbS
or InGaAs). Such a sensor would be a dynamic tool for process innovation
and continuous quality improvement using risk-based models for inspection.
The authors are grateful for the partial support from both the
Kentucky Science and Education Foundation (KSEF-148-502-03-61) and
the National Institutes of Health (N01AA 33003). The authors would also
like to acknowledge the contributions of Jim Shanley and Jay Oliver of Isis
Pharmaceuticals for their assistance in feasibility studies and Quality
Assurance logistics.
1. Lodder R, Hieftje G. App. Spec. 1988, 42, 556-558.
2. Lodder R, Selby M, Hieftje G. Anal. Chem. 1987, 59, 1921-1930.
3. Lodder R, Hieftje G. App. Spec. 1988, 42, 1500-1512.
4. Drennen JK, Lodder RA. J. Pharm. Sci. 1990, 79, 622-627.
5. Vergote GJ, Vervaet C, Remon JP, Haemers T, Verpoort F. Eur. J. Pharm.
Sci. 2002, 16(1,2), 63-67.
6. Burns D, Ciurczak E. Handbook of Near-Infrared Analysis (Practical
Spectroscopy) – 2nd edition. Marcel Dekker, 2001.
7. Jolliffe IT. Principal Component Analysis. New York: Springer, 2002.
8. Hamilton S, Lodder R. Proceedings of Society of Photo-Optical
Instrumentation Engineers 2002, 4626, 136-147.
9. Geladi P, MacDougall D, Martens H. App. Spec. 1985, 39, 491.
10. Maddams W, Southon M. Spectrochim. Acta. 1982, 38A, 459-466.
11. Drennen JK, Lodder RA. Spectroscopy. 1991, 6(8), 34-39.
Joseph Medendorp received his B.A. in Chemistry from Cornell
University in 2002, and is currently a 4th year Ph.D. student in
Pharmaceutical
Sciences at the University of Kentucky. He is currently a UK-Center for
Pharmaceutical Science and Technology Graduate Research Fellow.
Joseph Wyse, Ph.D., is the General Manager of the University of
Kentucky Center for Pharmaceutical Science and Technology. Dr. Wyse has
fourteen
years of product and business development experience in the US
pharmaceutical and biotech industry. His past positions include serving
a
group leader at Cryopharm Corporation (Pasadena, CA), manager at
LifeCell Corporation (The Woodlands, TX) and associate director at
Aronex
Pharmaceuticals Inc. (The Woodlands, TX).
Robert A. Lodder, Ph.D. is currently Professor of
Pharmaceutical Sciences at the College of Pharmacy, University of
Kentucky Medical Center. Dr.
Lodder holds joint appointments as a professor in the Departments of
Chemistry and Electrical/Computer Engineering. He serves on the board
of
directors of Spherix and is a member of the U.S. Food and Drug
Administration Advisory Committee on Pharmaceutical Science, Process
Analytical
Technologies subcommittee.
Lloyd Tillman, Ph.D., is Executive Director of Pharmaceutical
Development at Isis Pharmaceuticals, Inc., Carlsbad, CA, where he is
responsible
for the formulation research, development and clinical manufacture of
antisense oligonucleotide drug products. Prior to joining Isis, Dr.
Tillman
worked at the FDA from 1994 to 1997 as Associate Director over the
Product Quality Research Laboratory.
Sujatha Sonti, Ph.D., is currently Associate Director,
Technical Affairs at SkinMedica, Carlsbad, CA, where she is responsible
for product development
of topical formulations. Prior to joining SkinMedica, Dr. Sonti was
responsible for formulation development and clinical manufacturing of
oral
and topical dosage forms at Isis Pharmaceuticals, Inc., Carlsbad, CA.
Michael Jay, Ph.D., is a Professor of Pharmaceutical Sciences
in the College of Pharmacy at the University of Kentucky and the
Director of the
Center for Pharmaceutical Science & Technology. The Center has
facilities for analytical method development and validation, stability
studies, formulation
development, and is home to an FDA registered pharmaceutical cGMP
manufacturing facility.
To correspond with the authors, please contact the editor at: ay@russpub.com
*This article was published in the May/June 2006 issue of PAT, Volume 3, Issue 3, on pages 11-17.